Parameter
Code No.
Equivalent
Entry
Defaults/
Start-Up
Value
Unit Customer
Entered
Value
Description Enterable Range
of Values
Recommended
Range
* Cannot be user set.
§ Is dependent upon codes 11 and 18.
‡ Parameter 11 plus Parameter 18 must not exceed 199.99.
67 0.00 * Corrupt memory
68 0.68 Loop gain check request - set to zero to initiate test 0 to initiate test
69 0.00 % O
2
Loop gain error
70 Cal 0 0.08 or 0.70 Calibrate pressure compensation (if fitted). Set to zero 0 to calibrate pressure
71
72
} Used for setting security codes - not displayable in
display mode
75 2.00 % O
2
Zero gas tolerance limits 0 to 5.00 0.0 to 2.00
77 Cal Zero 0.77 Calibrate zero point - set to zero 0 to calibrate zero
78 Cal Span 0.78 Calibrate span point - set to zero 0 to calibrate span
85 25.00 % O
2
Span gas tolerance limits 0 to 50.00 0 to 50.0
87 Valve zero 0.87 Zero gas solenoid valve 0 to select zero gas
88 Valve span 0.88 Span gas solenoid valve 0 to select span gas
89 Valve sample 0.01 Sample gas solenoid valve 0 to select sample gas
92 57.00 °C * Cell temperature low alarm
93 63.00 °C * Cell temperature high alarm
94 3.00 * Cell temperature alarm deviation from set point
95
97 Gas Zero
0.67
0.00
* Alarm parameter store
Calibration zero point -10.00 to 21.00 -10.00 to 21.00 param
98 Gas span 100.00 Calibration span point 0 to 163.00 0 to +100.00
99 Gas sample variable % O
2
* Oxygen reading of sample gas