EasyManua.ls Logo

SICK DUSTHUNTER T - Page 186

SICK DUSTHUNTER T
552 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Clearing Malfunctions
GMS800 · Operating Instructions · 8013025 V1.2 · © SICK MAIHAK GmbH 59
Subject to change without notice
9.3 If measured values are obviously incorrect …
9.4 If measured values are unstable for no apparent reason …
Possible causes Information Notes for service
GMS800 not ready for opera-
tion.
b
Check operational readiness
(
p. 36, §5.3).
GMS800 not measuring the
sample gas.
b
Check sample gas path and all
valves (e.g. switching from test gas to
sample gas).
b
Make sure valves are function-
ing correctly, disassemble if
necessary.
Sample gas path not switched
correctly.
GMS800 not correctly
adjusted.
b
Check prerequisites for correct
adjustment:
Correct test gases used?
Nominal values set correctly?
b
Then run an adjustment.
b
Check test gases used critically
(nominal values, manufacturing
tolerance, state).
Measuring parameter settings
not suitable for the applica-
tion.
b
Check corresponding settings (e.g.
damping). try a different setting as
test.
Sample gas pressure inside
the GMS800 too high.
b
Ensure sample gas pressure in the
GMS800 is not greater than 20 kPa
(= 200 mbar) against atmospheric
pressure.
The gas pressure can influence
measured values in most of the
measuring principles used.
Sample gas path not gas-
tight.
b
Visually inspect the installation.
b
When a defect is suspected: Inform
the manufacturer's Customer Ser-
vice or trained skilled persons.
Carry out a leak tightness check.
When only observed on one
measured value output: the
load is too high.
b
Make sure that the total internal
resistance of the connected devices
is not larger than 500 .
b
Measure including feed line.
Analyzer module is contami-
nated.
b
Contact manufacturer’s Customer
Service or trained skilled persons.
b
Inspect the measuring cell/
cuvette.
b
Clean or replace if necessary.
With computation of an ana-
log input (option): External
analog signal is erroneous or
has failed.
b
Check external equipment that deliv-
ers the analog signal for cross-sensi-
tivity compensation.
Connection interrupted?
Problem with external measure-
ment?
External analyzer not adjusted?
Possible causes Information Notes for service
Strong pressure fluctuations
at the sample gas outlet.
b
Install separate exhaust gas line for
the GMS800 .
Strong mechanical vibrations.
b
Check ambient conditions where the
GMS800 is installed.

Table of Contents

Related product manuals