FLOWSIC100 Operating Instructions
Gas Velocity Monitor
Assembly and Installation
78
©
SICK MAIHAK GmbH · Germany · All rights reserved
88012513/03-2008
3.3.3 Installing and Connecting the Sender/Receiver Units
Check the following before installation:
The sender/receiver units must have at least the same nominal length as the flanges
with pipe.
The inside of the flange pipes must be free of welding beads.
The inside of the probe pipes on the sender/receiver units must not come into contact
with the flange pipes.
The cable connection on the electronics unit for sender/receiver units with digital signal
transmission must be at the bottom.
Note: If this does not comply with the installation requirements of the type FLSE100-PR as
shown in Fig. 3.4, loosen the screwed connections between the electronics unit and PR
/ PRAC connection, rotate the unit accordingly (90
°, 180°, 270°), and screw the
components together again.
Purge-air Supply for Internally Cooled Sender/Receiver Units FLSE100-MAC/HAC
The purge-air unit must be in operation.
Connect the purge-air hose DN25 to the purge-air connection on the sender/receiver
unit (mount the hose clamp on the free hose end, connect the purge-air hose and secure
it with the hose clamp).
Make sure, that the purge-air flows in and out from below (see Fig. 3.22).
If this is not guaranteed, loosen the screwed connections between the purge-air connec-
tion and the duct probe, rotate the unit accordingly (90
°, 180°, 270°), and screw the
components together again
If the optional purge-air supply in the connection box is used, slide the lose end of the
purge-air hose on adapter 40-25 and secure it with the hose clamp.
Purge-air Supply for Internally Cooled Sender/Receiver Unit FLSE100-PRAC
The sender/receiver unit must be installed so that cable connection and air intake of the
pump are oriented downwards.
It has to be ensured that only contamination-free and dry air is taken in (use weather-
proof covers if necessary).