EasyManua.ls Logo

SICK DUSTHUNTER T - Page 271

SICK DUSTHUNTER T
552 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions FLOWSIC100
Gas Velocity Monitor
Assembly and Installation
88012513/03-2008 © SICK MAIHAK GmbH · Germany · All rights reserved 79
Purge-air Supply for Purged sender/receiver units
If used with corrosive gases, the sender/receiver units with the next longest nominal
length than that of the flanges with pipe must be used to minimize corrosion (see Section
2.3.1.1).
The purge-air unit must be in operation.
Connect the purge-air hoses.
Mount the hose clamp loosely on the free hose end, connect the purge-air hose to the
purge-air connection on the sender/receiver units, and secure it with the hose clamp.
The purge-air must be supplied from below; the purge air must flow out in the direction
of the gas flow (see Fig. 3.22).
Fig. 3.22: Arrangement of cable connection and purge-air supply for internally cooled and purged
sender/receiver units (figure shows type FLSE100-PM/PH)
Installation and connection
DANGER
Warning
The sender/receiver units may only be mounted if it is safe to do so (for example, if the in-
stallation has been shut down, see Section 1.3.3).
Remove the blind plug from the flange.
Insert the sender/receiver units in the flanges with pipe as previously described, and
screw the components together.
Connect the cable to the evaluation unit at the connector on the sender/receiver unit.
Discharged
purge air
Gas flow
Cable connection
Purge-air supply
Gas
flow
Cable connection
Purge-air supply
Discharged
purge air
Ultrasonic
transducer
Duct
Figure shows type FLSE 100
PM / PH mounted on
vertical duct

Table of Contents

Related product manuals