GM35 Operating Instructions
Probe Model
Handling the evaluation unit
42 © SICK MAIHAK GmbH • Germany · All rights reserved 8009389/07-2006
Fig. 22 Menu structure of the GM35/part 2
Configuration
Components:
CO CO2 H2O
Probe: CAN
Physical Unit
Unit: ppm
Ref. conditions
Average
Avg. Time: 60 s
Meas. Distance
Active: 2000 mm
Tube 1:
Tube 2:
Temperature
Source: Probe
Subst.: 30 K
Unit: K
Pressure
Source: Probe
Subst.: 1013 hPa
Analog Out
Live Zero: mA
Output 1: CO
Output 2: CO2
Output 3: H2O
Analog In
Input 1: T
Input 2: p
Input 3: v
Check Cycle
Rep. Time: 120 min
Regress. Funct. CO
CO
CO2
H2O
Profibus
Address: 126
Parameterization
Measurement components:
CO, CO
2
, H
2
O
CAN connection (probe
or cross duct)
Unit and reference value
settings for the individual
gases
Averaging time setting for the
system
Measuring distance: Input
range 1…12000 mm
Purge air distances SR unit/ref.
side
Temperature setting: source
(probe/replacement/AI1) and
unit (K, °C, °F)
Pressure setting: source
(probe/replacement/AI2)
Analog output settings: live-
zero, comp. (CO, CO
2
, H
2
O,
O
2
, T, p - -) display range, cycle
output
Analog input settings: e.g. AI1
unit (K, °C, °F), live-zero (0, 2,
4 mA, read range)
Check cycle interval setting:
0…1440 min
Regression function comp.:
CO
2
/H
2
O
Span: 0.50…1.99,
Zero: 0…±9999.9
Option
PROFIBUS setting:
enter bus address.
Continued on page Seite 43