How to apply range of Flow rate
[Picture 25]
1) Reference
dp: Maximum range of Delta– P
VE: Maximum range of Velocity
FR: Maximum range of real-time Flow rate
1M: Maximum range of 1min Flow rate
5M: Maximum range of 5 min Flow rate
30: Maximum range of 30 min Flow rate
1H: Maximum range of 1hour Flow rate
2) Setting-up SPAN RANGE
! Caution: To set-up accurate dP and VE of SPAN RANGE, accurate data of Stack
should be typed in‘ F2: System Setting-up’ .
dP: Stack cross section, Pitot factor, Humid flue gas density.
VE: Stack cross section
3) How to modify Setup
â‘ Move to item to modify and press Enter key.
â‘¡ Modify set-up value with CURSOR(Up/Down/Left/Right) key.
â‘¢ If end all, press Enter key.