Start-up and Parameter Settings
DUSTHUNTER T · Operating Instructions · 8012428 V 2.0 · © SICK MAIHAK GmbH Germany 67
4.4.3 Determining the check cycle
Interval time, control value output on the analog output and the starting timepoint for
automatic check cycles can be modified in the "Adjustment / Function Check - Automatic“
directory.
Figure 52 "Adjustment / Function Check - Automatic“ directory (example for settings)
Default values
→
p. 66, §4.4.2
Entry field Parameter Remark
Output duration of
function control
value
Value in seconds Output duration of check values
Output control
values at AO
Inactive Control value is not output on the analog output
Active Check cycle is output on the analog output.
Function check
interval
Time between two
check cycles
→
p. 16, §2.1.3
Function Check
Start Time
Hours Defining a start timepoint in hours, minutes and seconds
Minutes
Seconds
The value measured last is output during control value determination (
→
p. 16,
Fig. 3).