ID Icons Message Cause/Action
102 Volume flow below warn‐
ing limit
Check process conditions. Adjust the parameter value
'Lower warning limit' to normal process conditions.
103 Volume flow below alarm
limit
Check process conditions. Adjust the parameter value
'Lower alarm limit' to normal process conditions.
104 Density above alarm limit Check process conditions. Adjust the parameter value 'Up‐
per alarm limit' to normal process conditions.
105 Density above warning
limit
Check process conditions. Adjust the parameter value 'Up‐
per warning limit' to normal process conditions.
106 Density below warning
limit
Check process conditions. Adjust the parameter value
'Lower warning limit' to normal process conditions.
107 Density below alarm limit Check process conditions. Adjust the parameter value
'Lower alarm limit' to normal process conditions.
108 Medium temperature
above alarm limit
Check process conditions. Adjust the parameter value 'Up‐
per alarm limit' to normal process conditions.
109 Medium temperature
above warning limit
Check process conditions. Adjust the parameter value 'Up‐
per warning limit' to normal process conditions.
110 Medium temperature be‐
low warning limit
Check process conditions. Adjust the parameter value
'Lower warning limit' to normal process conditions.
Diagnostics and troubleshooting
10.2 Fault codes and corrective actions
SITRANS FC330
Operating Instructions, 01/2019, A5E44030648-AB 177