EasyManua.ls Logo

Siemens WEBUI - Figure 4-3 Calibration Required; Figure 4-4 Island Calibration; Island Calibration

Siemens WEBUI
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CALIBRATION & ADJUSTMENT
4-12
SIG-QG-20-02 JANUARY 2021
Version : A.2
If the calibrations are required, these are marked with a red check box as shown in Figure 4-3.
Figure 4-3 Calibration Required
To select which calibration to perform, either click on a specific calibration in the menu bar, for example,
GCP shown in the yellow box above, or select the calibration from the drop-down menu shown in green
box (Figure 4-3).
4.1 ISLAND CALIBRATION
When the island calibration is selected, the WebUI will show the distance at which the hard wire shunt
should be placed. The distances shown are automatically calculated for the island frequency that has
been configured.
Figure 4-4 Island Calibration

Table of Contents

Related product manuals