15
SKOPE TME/SKT ActiveCore
Electronic Controller
Service Manual
Hardware Setup
Hardware
Inputs
The controller has three hardware inputs as detailed in the table below. All use pin 9 as
common.
Temperature
Settings
The standard temperature settings for this fridge are set up for perishable product. All product
temperatures are continuously below 5°C in Normal and Energy Saving mode. The Normal
mode temperature can be altered by changing the setpoint (parameter St), and the Energy
Saving mode temperature altered by changing the Energy Saving mode setpoint delta
(parameter r4).
Temperature
Probe
Readings
Control probe
If parameter /4 is set to 1, the control probe temperature reading is displayed on the controller
faceplate during normal operation. If parameter /4 is not set to 1, it can be changed to 1 to view
the control probe temperature reading.
Evaporator probe
To view the evaporator probe temperature reading, press and hold the Energy Save (up) and
Set (mute) buttons. The temperature reading will display on the controller faceplate.
Wiring
Termination
Refer to the diagram below for controller termination details.
Programming the Electronic Controller
The electronic controller parameter configuration program is set by SKOPE at the factory. A
label on the controller box indicates the parameter configuration program number (e.g. the
ActiveCore Series use program ‘081’).
The electronic controller parameters can be modified using the keypad. Frequent parameters
can be access and changed without entering a password (e.g. change normal and energy save
Table 5: Electronic controller hardware inputs
Pins (on rear of controller) Hardware description
9-8 Control probe
9-10 Evaporator probe
9-11 Door switch
Door switch (11)
Evaporator probe (10)
Common (9)
Control probe (8)
Supply (6, 7)
Compressor/fans (2, 4, 5)
Lights (1)

Table of Contents