EasyManua.ls Logo

Skope TME ActiveCore - Page 72

Skope TME ActiveCore
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72
Troubleshooting
Service Manual
SKOPE TME/SKT ActiveCore
Moisture build up on
door or exterior.
High humidity.
Check ambient operating temperature and reposition fridge
if necessary.
Frequent door opening. Limit door openings.
Door not closing properly. Check and clean door gasket.
Fridge door does not
shut properly.
Fridge is on an uneven surface. Level the fridge.
Door is obstructed. Check shelves and product.
Warm cabinet
temperatures
Compressor operating
for long periods (more
than 1 hour)
Blocked condenser Clean the condenser filter (if fitted) and condenser.
Poor ventilation around
refrigeration unit
Ensure the cabinet has good ventilation around the
refrigeration unit.
Ensure the cabinet is within the maximum operating
temperature.
Table 35: Cabinet and unit troubleshooting (continued)
Problem Possible cause Repair