6-5
Stage Error Message Error Handling
Assay Parameter Version
Mismatch (Error Code 108)
Please contact the authorized
technical support department to get
the latest version of firmware or assay
parameter.
Used Disc
(Error Code 201)
Please remove the used reagent disc
and insert a new reagent disc.
Insufficient Sample
(Error Code 204, 213)
The sample volume is insufficient.
Please insert a new reagent disc and
make sure to inject 20L of sample.
Disc Error
(Error Code 202, 203, 211,
212,323,355,841,842,843)
Please insert a new reagent disc.
Device Abnormal
(Error Code 1~20, 30, 60~62,
3XX, 800, 801, 802)
Please remove the reagent disc,
reboot the anaylzer, and insert a new
regent disc. If the situation is not
resolved, please contact the
authorized technical support
department.
Settings
Password Error Please enter the correct password.
Date & Time Error Please enter the correct date and time.
USB Drive Error
(Error Code 411)
Please check that the USB drive is
properly plugged into the device.
Please check whether the USB drive
space is full. If the situation is not
resolved, please contact the
authorized technical support
department.
No Software Found There is no Software update file
found within the USB drive. Please
contact the authorized technical
support department for this firmware.
Upgrade Failure (Error Code
950, 951, 953, 959)
Please contact the authorized
technical support department.
Assay Parameter Upgrade
Failure (Error Code 950, 953)
Please contact the authorized
technical support department.
Internal Memory Full
Please contact the authorized
technical support department.
Invalid ID Please enter the correct Operator ID.