A-13
> (0x3E) - The analysis result is greater than the upper limit of the
dynamic range.
Byte 14 to 20: Normally filled with the analysis result. If Byte 13 is <, it is below the
lower limit of dynamic range. If Byte 13 is >, it is above the upper limit
of dynamic range. If “N.A.” is displayed, the analysis result cannot be
calculated. If test results are influenced by more than 20% due to LIP,
HEM, or ICT, L: _: _, _: H: _, _: _: I, L: H: _, L: _: I, _: H: I, or L: H: I
will be displayed. L indicates LIP interference, H indicates HEM
interference, and I indicates ICT interference.
Byte 22 to 35: Reference range.
Byte 37 to 51: Unit symbol.
Note: If males and females are distinguished for the normal value range, (19) will be
output twice. However, only “Bytes 22-35: Reference range” and “Bytes 37-51:
Unit symbols” will be output the second time, while the rest of the byte contents
will be left blank.
(20) is as described in (19), but the times sent are “Total item number - 1” times.