Shenzhen Skyshl Technology Co.,LTD.
Sales email: sales1@skyshl.net After Sales email:af@skyshl.net Website:www.skyshl.net
SS413F Optical Fiber Fusion Splicer Userās Manual
V
Setting fiber alignment method.
"Precise":The fiber is aligned according to the the fiber core
position and the cladding position at the same time.
"Fiber core": The fiber is aligned according the the fiber core
position.
"Cladding": The fiber is aligned according the the cladding
position.
If "tension test" is set to "open", after the splicing is completed,
the tensile test is performed automatically.
Setting Cleave angle limit. When the Cleave angle of any of the
fibers exceeds the limit, a limit alarm message appears on the
screen.
When the estimated splice loss exceeds the setting splice loss
limit, a limit alarm message appears on the screen.
Time for
cleaning
discharge
Cleaning discharge is a short duration arc and used to remove
tiny dust on the surface of the fiber. Changing this parameter can
change the length of the cleaning discharge time.
Manually
adjusting arc
time
Under certain conditions, the splice loss can be improved by
more than one time of arc. Changing this parameter can change
the length of the arc time.