3.3.1
ENABLE
The
sorting method.
This
may
be
chosen
from:
*
[
off
]
No sorting
done.
[
continuous
]
Every
component
sorted.
[fixed
count]
Every
nth
component
sorted.
(n
=
STEP
SIZE.)
[random
]
Random
sort,
within
a maximum
step
size.
9-3.2 ITEM
The
display
parameter
sorted.
*
[
par
l
1
Parameter
1 sorted.
*
[
par
2
j
Parameter
2 sorted.
On entry
of the
display
parameter
setting
you are
invited
to enter
the
parameter
value.
This is the
nominal
value
to which
the tolerance
values,
specified under BINS,
refer.
9.3.3
BINS
First you specify
the
number
of bins
into which
the
components
are
to be
sorted Then,
for
each bin,
you
define
the
MIN% and
MAX%
tolerance
values,
each of which
refers
to the
parameter
value
entered under
ITEM.
A
physical set
of component
bins
may
be
used to
store the
components
as
their
corresponding
bin
number
is
displayed.
9.3.4
STOP
Defined
conditions for
stopping
test.
[off]
Test
stopped only
when
ENABLE
[off]
is selected.
[on]
Measurements
stopped
after
a number
of tests
have
been made.
This
number
is defined
by the user.
9.4
[BINSORT
B]
The
[BINSORT
B] menu
allows
you to
choose
the
parameters
for
operating
a
component
test machine.
These are:
the number
of
attempts to
be made
to
get a
"pass” result.
the machine
drive levels,
the machine
drive logic.
9.4.1
RETRY
Enter
the maximum
number
of times
each
component
is
to be
measured
in
an
attempt
to obtain
a pass result.
5.34
JWS/1
260/2