iBED Awareness System Error Codes
Error
Code
Name Description Type Affected Systems
900 Not Used Message Errors
- When the timeout for the message error has
expired the display should return to the status
screen unless otherwise noted or unless a
higher priority screen is being handled.
901 ScaleSystem When turning on the Scale system, using the
Scale Button, if a “Logged” error is present
that affects the scale system (see affected
systems column) the message “Function
Disabled - Call Service” will blink.
System N/A
902 BedExitSystem When turning on the Bed Exit system if a
“Logged” error is present that affects the
scale or bed exit systems (see affected
systems column) the message “Function
Disabled - Call Service” will blink.
Bed Exit will alarm if Bed Exit is armed and a
200 level exception occurs. The message will
be displayed “Turn off Bed Exit - Call Service”.
When the system is disarmed, the display will
go back to the status screen. Since the bed
exit alarm has the highest priority, if the Bed
Exit alarm is tripped due to the error condition
this message will not be seen.
Message N/A
903 LBSSystem When turning on the iBED Awareness system
if a “Logged” error is present that affects
the iBED Awareness system (see affected
systems column) the message “Function
Disabled - Call Service” will blink.
iBED Awareness will alarm if iBED Awareness
is “On” and a 200 or 300 level Logged error
occurs. The message will be displayed “Turn
off Bed Status - Call Service”, When the
system is turned off, the display will go back
to the status screen.
Message N/A
904 UnknownBedType Unknown bed type occurs when the
configuration switch bank does not contain
a value that is recognized by the software.
This will cause the product to stop all function
and product a message screen indicating the
error number and the message “Unknown
Bed Type - Call Service”
Message N/A
ERROR MESSAGES (CONTINUED)
NOTE: The 900 error codes are not logged.