In case of <profile> = 1, 2 or 3 the instrument sends results in the format defined as follows:
#2 [,<aver>],<profile>,Xc,(...);
where c is the value of the result X or question mark (?) if result X is not available;
If no results are available the instrument will send:
#2,?;
The codes of the results from the SLM mode are defined as follows:
v under-range flag (ccc equals to 0 when the overload did not occur, 2 when the under-range took
place during the last measurement period but did not occur in the last second of the measurement
and 3 when the under-range took place during the last measurement period and it lasted in the last
second of the measurement);
V overload flag (ccc equals to 0 or 1);
T time of the measurement (ccc – value in seconds);
x start date of the measurement in format dd/mm/yyyy (dd – day, mm – month, yyyy - year)
t start time of the measurement in format hh/mm/ss (hh – hour, mm – minute, ss - second)
P Lpeak value (ccc – the value in dB);
M Lmax value (ccc – the value in dB);
N Lmin value (ccc – the value in dB);
S L result (ccc – the value in dB);
R Leq result (ccc – the value in dB).
U LE result (ccc – the value in dB);
B(k) Lden result (ccc – the value in dB; k – flag determining the kind of the result);
I(nn) LEPd result (ccc – the value in dB, nn – the value of Exposure Time in minutes);
Y Ltm3 result (ccc – the value in dB);
Z Ltm5 result (ccc – the value in dB);
L(nn) value L of the nn statistics (ccc – the value in dB).
g LR15 result (ccc – the value in dB;);
G LR60 result (ccc – the value in dB;).
Note: In the case of Lden, the value k placed in the parenthesis after the code B, denotes the
kind of the currently measured result. The kind of the Lden result depends on the time during
which the measurements were performed (d denotes day, e denotes evening and n denotes
night). The corresponding values of k parameter and the kind of the measured Lden result are
presented below:
k = 1 Ld result,
k = 2 Le result,
k = 3 Lde result,
k = 4 Ln result,
k = 5 Lnd result,
k = 6 Len result,
k = 7 Lden result.
The codes of the results from the DOSE METER mode are defined as follows:
v under-range flag (ccc equals to 0 when the overload did not occur, 2 when the under-range took
place during the last measurement period but did not occur in the last second of the measurement
and 3 when the under-range took place during the last measurement period and it lasted in the last
second of the measurement);