V overload flag (ccc equals to 0 or 1);
T time of the measurement (ccc – value in seconds);
x start date of the measurement in format dd/mm/yyyy (dd – day, mm – month, yyyy - year)
t start time of the measurement in format hh/mm/ss (hh – hour, mm – minute, ss - second)
P Lpeak value (ccc – the value in dB);
M Lmax value (ccc – the value in dB);
N Lmin value (ccc – the value in dB);
S L result (ccc – the value in dB);
D DOSE result (ccc – the value in %);
d D_8h result (ccc – the value in %);
p PrDOSE result (ccc – the value in %);
A LAV result (ccc – the value in dB);
R Leq result (ccc – the value in dB);
U LE result (ccc – the value in dB);
u SEL8 result (ccc – the value in dB);
E E result (ccc – the value in Pa
2
h);
e E_8h result (ccc – the value in Pa
2
h);
I(nn) LEPd result (ccc – the value in dB, nn – the value of Exposure Time in minutes);
J PSEL result (ccc – the value in dB);
Y Ltm3 result (ccc – the value in dB);
Z Ltm5 result (ccc – the value in dB);
L(nn) value L of the nn statistics (ccc – the value in dB);
C PTC result (ccc – the counter value);
c PTP result (ccc – the value in %);
l ULT result (ccc – value in seconds);
W TWA result (ccc – the value in dB);
w PrTWA result (ccc – the value in dB);
a Lc-a result (ccc – the value in dB);
The exemplary results of the instrument’s response after sending to it the following sequence
of characters: #2,1; coming from the first profile are given below:
a) for the case of the SLM mode:
#2,1,x17/03/2014,t13:44:28,v0,V0,T10,P79.97,M52.92,N38.50,S46.35,R43.91,U53.91,B(1)43.91,I(
480)43.92,Y50.67,Z51.15,L(01)55.00,L(10)45.60,L(20)44.30,L(30)42.80,L(40)41.50,L(50)40.80,L(6
0)40.40,L(70)40.00,L(80)39.50,L(90)39.00;
b) and for the case of the SEM mode:
#2,1,x17/03/2014,t13:48:36,v0,V0,T7,P124.39,M99.26,N41.54,S42.05,D0,d389,p389,A85.86,R85.
86,U94.31,u130.45,E0.00,e1.23,I(480)85.87,J49.72,Y95.62,Z99.22,L(01)100.30,L(10)89.50,L(20)7
8.60,L(30)68.50,L(40)60.30,L(50)54.00,L(60)51.00,L(70)46.50,L(80)44.00,L(90)42.40,C4,c6,l0,W4
9.72,w85.87,a-0.55;