6-10
Model 3550 LCR Meter Instruction Manual Service Information
Model 3550 Verification Procedure cont’d:
TEGAM 3550 AND AGILENT 4284A MEASUREMENTS
A B A-B
STANDARD
VALUE
FREQ
(Hz)
LEVEL
(V)
PARAMETER 4284A 3550 ERROR TOL.
(CTS)
100mƻ 1k 1 Z
=(1000A-B)*100
1400
100mƻ 1k 1 LJ =(A-B) *100 101
1ƻ 1k 1 Z =(A-B) *10k 340
1ƻ 1k 1 LJ =(A-B) *100 20
10ƻ 1k 1 Z =(A-B) *1k 94
10ƻ 1k 1 LJ =(A-B) *100 10
100ƻ 1k 1 Z =(A-B) *100 22
100ƻ 1k 1 LJ =(A-B) *100 10
1kƻ 1k 1 Z =(A-B) *10k 21
1kƻ 1k 1 LJ =(A-B) *100 10
10kƻ 1k 1 Z =(A-B) *1k 21
10kƻ 1k 1 LJ =(A-B) *100 10
100kƻ 1k 1 Z =(A-B) *100 21
100kƻ 1k 1 LJ =(A-B) *100 10
100ƻ 42 1 Z =(A-B) *100 27
100ƻ 42 1 LJ =(A-B) *100 10
1kƻ 42 1 Z =(A-B) *10k 26
1kƻ 42 1 LJ =(A-B) *100 10
100ƻ 10k 1 Z =(A-B) *100 27
100ƻ 10k 1 LJ =(A-B) *100 10
1kƻ 10k 1 Z =(A-B) *10k 27
1kƻ 10k 1 LJ =(A-B) *100 10
100ƻ 100k 1 Z =(A-B) *100 85
100ƻ 100k 1 LJ =(A-B) *100 40
1kƻ 100k 1 Z =(A-B) *10k 85
1kƻ 100k 1 LJ =(A-B) *100 40
10ƻ 1M 1 Z =(A-B) *1k 204
10ƻ 1M 1 LJ =(A-B) *100 50
100ƻ 1M 1 Z =(A-B) *100 115
100ƻ 1M 1 LJ =(A-B) *100 40
1kƻ 1M 1 Z =(A-B) *10k 115
1kƻ 1M 1 LJ =(A-B) *100 40
.01µF 1k 1 Z =(A-B) *1k 22
.01µF 1k 1 LJ =(A-B) *100 20
.1µF 1k 1 Z =(A-B) *10k 22
.1µF 1k 1 LJ =(A-B) *100 20
1µF 1k 1 Z =(A-B) *100 21
1µF 1k 1 LJ =(A-B) *100 20