Getting Started
Connecting t he instrument
to your PC(s)
You can use one o
f the following two methods to connect the instrument to your
PC(s):
If you are conn
ecting the instrument directly to a single PC, use an Ethernet
cable to connect between the Ethernet port on the instrument and the Ethernet
port on the PC.
If you are connecting the instrument to your local Ethernet network, use an
Ethernet cable to connect between the Ethernet port on the instrument and
the Ethernet hub port of your local network. By connecting to an Ethernet
network, you can access the instrument using any PC on the network.
To configure the network
parameters
The following two procedures describe how to configure the ECO8000 Series
network parameters. Use the first procedure if your network supports DHCP; use
the second procedure if your network does not support D HCP.
To config
ure parameters for a network with a DHCP server. If your network has
a DHCP server, perform the following steps to configure the instrument to
function as a DHCP client. When the instrument is connected to the network and
DHCP service is enabled, the instrument obtains the necessary network addresses
automatically from the DHCP server.
NOTE. Under some network environments, the instrument may not be able to
get t
he IP address automatically from a DHCP server. In this case, you need to
manually enter the appropriate network parameter values.
Ref
er to your network administrator or to the user documentation supplied with
your network server operating system (OS) for detailed information about DHCP
server functions.
1. Press the front-panel BACK button as necessary to access the top-level Main
menu.
2. Press the up (▲)ordown(▼) arrow button to select SYSTEM CONFIG,
andthenpresstheENTER button to access the SYSTEM CONFIG menu.
3. Press the up (▲)ordown(▼)arrowbuttontoselectNETWORK CONFIG,
and then p ress the ENTER button to access the NETWORK CONFIG
submenu. The display shows whether DH CP is enabled or not.
4. If necessary, press the left (◄)orright(►) arrow button to select Enable,
and then press the ENTER button to implement the selection.
5. Press the down (▼)arrowbuttontoselectIP ADDRESS.
8 ECO8000 Series Automatic Changeover Unit User Manual