EMAIL REPORTIN
G. Select this menu item and press the ENTER button to access
the EMAIL REPORTING submenu, which you can use to congure the e-mail
login name, password, and to/from addresses, and to congure which events
will trigger an e-mail to be sent from the instrument. (See page 76, EMAIL
REPORTING submenu.)
INTERNAL CL
OCK. Select this menu item and press the ENTER button to access
the INTERNAL CLOCK submenu, which you can use to set the local time of the
instrument and to congure the Daylight Savings Time (DST) schedule. (See
page 79, INTERNAL CLOCK submenu.)
FRONT PANEL TIMEOUT. Select this menu item and press the left ()orright
() arrow
button to set the front panel time-out period to either 1 minute or 5
minutes. Press the ENTER button to implement the selection. When the front
panel of the instrument has not been used for the amount of time congured as the
time-out period, the instrument control buttons become inoperative. In this case,
press and hold the PANEL ENABLE button until the instrument beeps.
The c ontrol buttons include the FAULT RESET, AUTO MODE, MANUAL
MODE, PRIMARY SOURCE, and BACKUP SOURCE buttons. The navigation
arrow buttons and the ENTER and BACK buttons are always operational and
are no
t affected by the time-out setting.
The backlight of the LCD display turns off after 5 minutes of inactivity to preserve
the l
ife of the backlight. When the instrument is powered on and the LCD display
is off, press any of the menu navigation buttons to turn on the display backlight.
LCD CONTRAST. Select this menu item and press the left ()orright() arrow
button to adjust the contrast of the LCD display. You can adjust the contrast
setting between 0 and 8.
GPI EVENTS. Select this menu item and press the ENTER button to access the
GPI EVENTS submenu, which you can use to select which events will trigger a
GPI signal to be sent from the instrument using the GPIO port. Enabled events will
cause pin 10 of the port to go low. (See page 22, GPIO connector signal pinouts.)
(See page 83, EVENT submenus.) (See page 49, Fault and event descriptions.)
BEEPER EVENTS. Select this menu item and press the ENTER button to access
the BEEPER EVENTS submenu, which you can use to select which events will
trigger a beep sound from the instrument. (See page 83, EVENT submenus.) (See
page 49, Fault and event descriptions.)
ECO8000 Series Automatic Changeover Unit User Manual 69

Table of Contents

Other manuals for Tektronix ECO8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix ECO8000 Series and is the answer not in the manual?

Tektronix ECO8000 Series Specifications

General IconGeneral
BrandTektronix
ModelECO8000 Series
CategoryControl Unit
LanguageEnglish

Related product manuals