Commands Listed in Alphabetical Order
Syntax
CALibrate:INT
ERNal:STATus?
Related Commands
*CAL?
Returns
This query will return one of the following:
INIT indicates the instrument has not had internal signal path calibration run.
WARMUP indicates the instrument has not warmed up for the required time
to perform signal path calibration.
PASS indicates that the signal path calibration completed successfully.
FAIL indicates that the signal path calibration did not complete successfully.
TEMPDRIFT indicates that the instrument has not reached a state of stable
temperature adequate for performing signal path calibration.
RUNNING indicates that the signal path calibration is currently running.
Examples
CALIBRATE:INTERNAL:STATUS? might return
:CALIBRATE:INTERNAL:STATUS INIT, indicating that the current status of the
internal signal path calibration is that it has not been run.
CALibrate:PRObestate:CH<x>? (Query Only)
This query-only command returns the probe calibration status for the probe of the
selected channel, 1 through 4.
Group
Calibration
Syntax
CALibrate:PRObestate:CH<x> ?
Related Commands
CH<x>:PROBECal?
Returns
This query will return one of the following:
–1 Indicating the probe calibration failed for the selected channel.
0 Indicating the probe calibration passe d for the selected channel.
1 Indicating the probe c alibration is initialized for the selected channel.
2 Indicating that the probe calibration for the selected channel is running.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-105