Commands Listed in Alphabetical Order
Arguments
<NR3> specifies
the A pulse width trigger lower limit, in seconds.
Examples
TRIGGER:A:PULSE:WIDTH:LOWL IMIT 2.0E-6 sets the lower limit of the
width trigger to 2 µ s.
TRIGGER:A:PULSE:WIDTH:LOWL IMIT? might return
:TRIGGER:A:PULSE:WIDTH:LOW LIMIT 1.0000E-9, indicating that the lower
limit of the width trigger is set to 1 ns.
TRIGger:{A|B}:PULse:WIDth:POLarity
This command sets or queries the polarity for the width trigger. This command is
equivalent to selecting Width Setup from the Trig menu and selecting the Polarity.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:WIDth: POLarity {NEGAtive|POSITI Ve}
TRIGger:{A|B}:PULse:WIDth: POLarity?
Related Commands
TRIGger:{A|B}:PULse:WIDth?
Arguments
NEGAtive specifies a negative pulse.
POSITIVe specifies a positive pulse.
Examples
TRIGGER:A:PULSE:WIDTH:POLA RITY NEGATIVE sets the pulse polarity to
negative.
TRIGGER:A:PULSE:WIDTH:POLA RITY? might return
:TRIGGER:A:PULSE:WIDTH:POL ARITY POSITIVE, indicating that the pulse
polarity is set to positive.
TRIGger:{A|B}:PULse:WIDth:POLarity:CH<x>
This command sets or queries the polarity for the A or B pulse width trigger for
the channel.
This command is equivalent to selecting Width Setup from the Trig menu and then
selecting the pulse width trigger Polarity for the channel.
Group
Trigger
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-543