Commands Listed in Alphabetical Order
Examples
MEASUREMENT:M
EAS1:UNITS?
might return :MEASUREMENT:MEAS1:UNIts %,
indicating units for measurement 1 are set to percent.
MEASUrement:MEAS<x>:VALue? (Query Only)
This query-only command returns the value that has been calculated for the
measurement specified by <x>, which ranges from 1 through 8. This command
is equivalent to selecting Display Statistics from the Measure menu and then
choosing V
alue from the drop-down list to display all measurement values
on-screen.
NOTE. This is the displayed value in the on-screen d isplay. If measurement
statistics are enabled, a new value is calculated with every waveform. In addition,
this value is updated about every 1/3 second. If you are acquiring a long
acquisition record, the instrument may take longer to update.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:VALue?
Related Commands
MEASUrement:MEAS<x>:UNIts?, *ESR?, ALLEv?
Examples
MEASUREMENT:MEAS1:VALUE? might return :MEASUREMENT:MEAS1:V ALue
2.8740E-06
. If the measurement has an error or warning associated with it,
then an item is added to the error queue. The error can be checked for with the
*ESR? and ALLEv? comma nds.
MEASUrement:METHod
This command sets or queries the method used to calculate the 0% and 100%
reference level. This command is equivalent to selecting Reference Levels from
the Measure menu and then choosing the desired Determine Base, Top From
setting.
Group
Measurement
Syntax
MEASUrement:METHod {HIStogram|MEAN| MINMax}
MEASUrement:METHod?
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-401