Commands Listed in Alphabetical Order
Syntax
TRIGger:B:RES
ET:THReshold {ECL|TTL|<NR3>}
TRIGger:B:RESET:THReshold?
Related Commands
TRIGger:B:RESET:STATE
Arguments
ECL se ts the reset t hreshold to ECL voltage levels.
TTL sets the reset threshold to TTL voltage levels.
<NR3> sets the reset threshold in volts.
Examples
TRIGGER:B:RESET:THRESHOLD TTL sets the trigger reset threshold level to
TTL.
TRIGGER:B:RESET:THRESHOLD? might return
:TRIGGER:B:RESET:THRESHOLD TTL, indicating that the trigger reset
threshold level is set to TTL.
TRIGger:B:RESET:TIMEOut
This command sets or queries the reset timer for a sequential timeout trigger
reset. For example, if the timeout is set to 1 s, this additional reset time b egins
following an A trigger event and runs sequentially with any B Trigger Delay. If a
B trigger event is not found before it expires, the instrument begins looking for a
new A event trigger sequence.
Group
Trigger
Syntax
TRIGger:B:RESET:TIMEOut <NR3>
TRIGger:B:RESET:TIMEOut?
Related Commands
TRIGger:B:RESET:TYPe
Arguments
<NR3> specifies the reset time.
Examples
TRIGGER:B:RESET:TIMEOUT 1.0E-6 sets the trigger reset timeout to 1 s.
TRIGGER:B:RESET:TIMEOUT? might return :TRIGGER:B:RESET:TIM EOUT
2.0000E-09
, indicating that the trigger reset timeout is 2 ns.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-613