Commands Listed in Alphabetical Order
Examples
TRIGGER:A:PUL
SE:RUNT:THRESHOLD:HIGH 120E-3
sets the upper limit of
the pulse runt trigger to 120 mV.
TRIGGER:A:PU
LSE:RUNT:THRESHOLD:HIGH?
might return
:TRIGGER:A:PULSE:RUNT:THR ESHOLD:HIGH 1.1000E+00, indicating that
the upper limit of the pulse runt trigger is currently set to 1.1 V.
TRIGger:{A
|B}:PULse:RUNT:THReshold:LOW
This comma nd sets or queries the lower limit for the pulse runt trigger. This
command is equivalent to selecting Runt Setup from the Trig menu and then
setting t
he Lower Level voltage.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:RUNT: THReshold:LOW <NR3>
TRIGger:{A|B}:PULse:RUNT: THReshold:LOW?
Related Commands
TRIGger:{A|B}:PULse:RUNT:THReshold?
Arguments
<NR3> specifies the threshold value in volts.
Examples
TRIGGER:A:PULSE:RUNT:THRE SHOLD:LOW 50E-3 sets the lower limit of the
pulse runt trigger to 50 mV.
TRIGGER:A:PULSE:RUNT:THRE SHOLD:LOW? might return
:TRIGGER:A:PULSE:RUNT:THR ESHOLD:LOW 1.2000E-01,
indicating that the lower limit of the pulse runt trigger is set to 120 mV.
TRIGger:{A|B}:PULse:RUNT:WHEn
This command sets or queries the type o f pulse width the trigger checks for when
it detects a runt. This is equivalent to selecting R unt Setup from the Trig menu
and choosing the desired Trigger When setting from the drop-down list.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:RUNT: WHEn {OCCurs|WIDERthan}
TRIGger:{A|B}:PULse:RUNT: WHEn?
2-530 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual