Commands Listed in Alphabetical Order
Related Commands
TRIGger:{A|B}
:PULse:RUNT:WIDth
Arguments
OCCurs argument specifies a trigger event if a runt of any detectable width occurs.
WIDERthan specifies a trigger event if a runt greater than the specified width
occurs.
Examples
TRIGGER:A:PULSE:RUNT:WHEN WIDERTHAN sets the runt trigger to occur when
the instrument detects a runt in a pulse wider than the specified width.
TRIGGER:A:PULSE:RUNT:WHEN? might return
:TRIGGER:A:PULSE:RUNT:WHEN OCCURS, indicating that a runt trigger will
occur if the instrument detects a runt of any detectable width.
TRIGger:{A|B}:PULse:RUNT:WIDth
This command sets or queries the minimum width for an Pulse Runt trigger. This
command is equivalent to selecting Runt Setup from the Trig menu and then
setting the Width.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:RUNT:W IDth <NR3>
TRIGger:{A|B}:PULse:RUNT:W IDth?
Related Commands
TRIGger:{A|B}:PULse:RUNT:WHEn
Arguments
<NR3> specifi es the minimum width, in seconds.
Examples
TRIGGER:A:PULSE:RUNT:WIDTH 15E-6 sets the m inimum width of the pulse
runt trigger to 1 5 µs.
TRIGGER:A:PULSE:RUNT:WIDTH ? might return
:TRIGGER:A:PULSE:RUNT:WIDT H 2.0000E-09, indicating that the minimum
width of a pulse runt trigger is 2 ns.
TRIGger:{A|B}:PULse:SOUrce
This command sets o r queries the source for the pulse trigger. This source
parameter applies to all classes of p ulse triggers. This command is equivalent
to selecting Event Trigger Setup from the Trig menu, selecting the pulse type
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-531