Commands Listed in Alphabetical Order
MASK:COUNt:W
AVEFORMS? (Query Only)
This query-only command returns the number of waveforms that have been
acquired and processed during pass/fail testing.
Group
Mask
Syntax
MASK:COUNt:WAVEFORMS?
Related Commands
MASK:COUNt:TESTS?, MASK:COUNt:TOTal?, MASK :COUNt:
VIOLATIONS?
Returns
<NR1>
Examples
MASK:COUNT:WAVEFORMS? might return :MASK:COUNT:WAVEFORMS 1000,
showing that 1,000 waveforms have been acquired and processed.
MASK:DISplay
This command sets or queries whether or not definedmasksaredisplayedonthe
screen. This is useful for temporarily turning off user-defined masks without
deleting them. It is also useful for removing a standard mask from the screen, but
leaving it as the selected standard. Mask counting, mask testing, and mask autos et
are unavailable if the mask display is Off. The default is On.
Group
Mask
Syntax
MASK:DISplay {OFF|ON|<NR1>}
MASK:DISplay?
Arguments
ON shows the masks on the display. This is the default value.
OFF removes the masks from the display.
<NR1> = 0 removes the masks from the display; any other value shows the masks
on the display.
Examples
MASK:DISPLAY ON sets the dis
play to show the d efined masks.
MASK:DISPLAY? might return :MASK:DISPLAY 1, i ndicating that t he defined
masks on the screen are disp
layed.
2-304 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual