<NR1> =0disabl
es the spec ied zoom; any other value e nables the specied
zoom.
Examples
ZOOM:ZOOM1:STATE ON turns Zoom1 on.
ZOOM:ZOOM2:STATE? might return :ZO OM:ZOOM2:STATE 1, indicating that
Zoom2 is on.
ZOOm:ZOOM<x>:<wfm>:DISplay
This command sets or queries the display of the specied waveform for the
speciedzoom,wherexisanintegerfrom1to4representingthedesiredzoom
window. The waveform affected is determine by <wfm> which can vary from
CH1, CH2, CH3, CH4, Math1, Math2, Math3, Math4, Ref1, Ref2, Ref3, or Ref4.
Group
Zoom
Syntax
ZOOm:ZOOM<x>:<wfm>:DISpla y {<NR1>|OFF|ON}
ZOOm:ZOOM<x>:<wfm>:DISpla y?
Arguments
<NR1> A zero disables the specied zoom; any other value enables the specied
zoom.
OFF disables the specied zoom.
ON displays the specied zoom.
Examples
ZOOM:ZOOM1:CH1:DISPLAY ON displays the specied zoom waveform.
ZOOM:ZOOM1:CH1:DISPLAY? might return :ZOOM:ZOOM1:CH1:DISPLAY 0,
indicating the channel 1 zoom is disabled.
ZOOm:ZOOM<x>:<wfm>:HORizontal:POSition
This command sets or queries the horizontal position of the specied waveform
for the speciedzoom,wherexisanintegerfrom1to4representingthedesired
zoom window. The setting of the ZOOm:ZOOM<x>:SCROLLLock command
determines the waveforms that are affected.
Group
Zoom
2-660 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix MSO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO70000 Series and is the answer not in the manual?

Tektronix MSO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelMSO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals