Commands Listed in Alphabetical Order
TRIGGER:MULTI
SCOPE:OPTION?
might return
:TRIGGER:MULTISCOPE:OPTIO N AND, indicating that the hub and all
participating instruments are set to AND triggering.
TRIGger:MUL
Tiscope:LOGic
This command sets or queries the TekLink trigger configuration when the TekLink
network connection is H UB.
Group
TekLink
Syntax
TRIGger:MULTiscope:LOGic {AND|OR}
TRIGger:MULTiscope:LOGic?
Arguments
AND sets the hub and all participating instruments to AND triggering configuration.
OR sets the hub and all participating instruments to OR triggering configuration.
Examples
TRIGGER:MULTISCOPE:LOGIC AND sets the hub and all participating instruments
to AND triggering configuration.
TRIGGER:MULTISCOPE:LOGIC? might return :TRIGGER:MULTISCOPE:LOGIC
OR
, indicating that the all participating instruments are set to OR trigger
ing
configuration.
TRIGger:MULTiscope:ROLe
This command sets or queries the role of the instrument in the MultiScope trigger
network. PRODUCER means that the instrument will both contribute and use the
MultiScope trigger. If there is no hub, this m akes the local instrument the master.
CONSUMER means that the trigger in from the MultiScope network is used
(rather than an internal trigger). This applies to a network with or without a hub.
Group
TekLink
Syntax
TRIGger:MULTiscope:ROLe {PRODucer|CONSume r|NONe}
TRIGger:MULTiscope:ROLe?
Arguments
PRODucer sets the instrument to both c ontribute and use the MultiScope trigger.
CONSumer sets the instrument to use the MultiScope trigger.
2-622 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual