Commands Listed in Alphabetical Order
Examples
MASK:USER:PRE
SAMPBITS 4
sets the presample bit value of the displayed mask
to 4.
MASK:USER:PRESAMPBITS? might return :MASK:USER: PRESAMPBITS 0,
indicating that the presample bit value of the displayed mask is set to 0.
MASK:USER:RECOrdlength
This command s ets or returns the nominal record length for pulse mask testing.
The query form of this command returns the record length value of the displayed
mask. The set form of this command only affects the user mask, regardless of
the current (displayed) mask.
Group
Mask
Syntax
MASK:USER:RECOrdlength <NR1>
MASK:USER:RECOrdlength?
Arguments
<NR1> is an integer number that sets the record length.
Examples
MASK:USER:RECORDLENGTH 500 sets the re cord length value of the displayed
mask to 500 points.
MASK:USER:RECORDLENGTH? might return :MASK:USER:R ECORDLENGTH 0,
indicating that the record length value of the displayed m ask is set to 0.
MASK:USER:SEG<m> (No Query Form)
This command deletes the specified mask segment from the user mask, whether or
not the user mask is t
he current mask.
Group
Mask
Syntax
MASK:USER:SEG<m> DELEte
Arguments
<m> is an integer that specifies the user mask segment number to delete from
the user mask.
DELETE removes the specified segment from the ma sk.
Examples
MASK:USER:SEG3 DELETE deletes mask segment three from the user mask.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-339