Commands Listed in Alphabetical Order
You can also lim
it the High measurement (normally taken over the entire
waveform record) to a gated region on the waveform.
HITs (histogr
am hits) measures the number of points in or on the histogram box.
LOW measures the Low reference (0% level, sometimes called Baseline) of
a waveform.
MAXimum nds the maximum amplitude. This value is the most positive peak
voltage fou
nd. It is measured over the entire waveform or gated region. When
histogram is selected with the MEASUrement:METHod command, the maximum
measurement measures the voltage of the highest nonzero bin in vertical
histograms or the time of the right-most bin in horizontal histograms.
MEAN amplitude measurement nds the arithmetic mean over the entire waveform
or gated region. When histogram is selected with the MEASUrement:METHod
command, the mean measurement measures the average of all acquired points
within or on the histogram.
MEDian (histogram measurement) measures the middle point of the histogram
box. Half of all acquired points within or on the histogram box are less than this
value
and half are greater than this value.
MINImum nds the minimum amplitude. This value is typically the most negative
peak v
oltage. It is measured over the entire waveform or gated region. When
histogram is selected with the MEASUrement:METHod command, the minimum
measurement measures the lowest nonzero bin in vertical histograms or the time
of the left-most nonzero bin in the horizontal histograms.
NCROss (timing measurement) measures the time from the trigger point to the rst
falling edge of the waveform or gated region. The distance (time) is measured at
the middle reference amplitude point o f the signal.
NDUty (negative duty cycle) is the ratio of the negative pulse width to the signal
period, express ed as a percentage. The duty cycle is measured on the rst cycle in
the waveform or gated region.
Negative Duty Cycle = (Negative Width) / Period × 100%
NOVershoot (negative overshoot) nds the negative overshoot value over the
entire waveform or gated region.
Negative Overshoot = (Low - Minimum) / Amplitude × 100%)
NWIdth (negative width) measurement is the distance (time) between the middle
reference (default = 50%) amplitude points of a negative pulse. The measurement
is made on the rst pulse in the waveform or gated region.
PBASe measures the base value used in extinction ratio measurements.
PCROss (timing measurement) measures the time from the trigger point to the rst
positive edge of the waveform or gated region. The distance (time) is measured at
the middle reference amplitude point o f the signal.
2-398 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix MSO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO70000 Series and is the answer not in the manual?

Tektronix MSO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelMSO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals