EasyManuals Logo

Tektronix TekScope THS710A User Manual

Tektronix TekScope THS710A
231 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
MEAS
THS710A, THS720A, THS730A & THS720P User Manual
3–33
High-Low Setup. The TekScope instrument determines the 10%, 50%,
or 90% levels of the selected waveform and then uses them to
calculate the measurements. You can choose the method used to
determine these levels:
H Histogram sets the values statistically; it finds the most common
value either above or below the midpoint (depending on whether
it is defining the high or low reference level). Since this
statistical approach ignores short-term aberrations (overshoot,
ringing, noise), histogram is the best method for measuring
digital waveforms and pulses.
H Min-max uses the highest and lowest values of the waveform
record. This method is best for measuring waveforms that have
no large, flat portions at a common value, such as sine waves and
triangle waves.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TekScope THS710A and is the answer not in the manual?

Tektronix TekScope THS710A Specifications

General IconGeneral
BrandTektronix
ModelTekScope THS710A
CategoryTest Equipment
LanguageEnglish

Related product manuals