7-2
Routine Maintenance and Disposal
Follow the manufactuer’s insturctions for preparation, handling, and use of the
cleaning agent.
After cleaning, visually verify contamination is removed and inspect parts for any
visual damage. If contamination is still visible, repeat cleaning process.
Contact Terumo Cardiovascular Technical Support if any parts appear damaged.
Use proper disposal suggested for the cleaning agent used.
Note: To disable the touchscreen for cleaning, see Chapter 5, “Touchscreen Deactivation.
Observe the following precautions during cleaning:
Caution
Do not expose the ends of the optical bers in the BPM(s),
the optical probe face or the optical reference color chip to
chemicals such as organic solvents, acids, strong bases or
abrasives. These chemicals can cause degradation in the
performance of the optical components. Use only water with a
soft, lint-free cloth to clean these surfaces.
Do not use harsh cleaning solutions on the cable connectors.
These can cause damage to the nish or the integrity of the
surfaces.
Do not use abrasive materials in cleaning the display screen. This
can cause degradation of the touchscreen functionality.
Note: Check the calibration uid residue in the calibrator pockets and on the BPMs
themselves.
Note: Do not submerge any part of this device in any liquid.
Battery Maintenance
A battery replacement message will appear at each startup when the battery is no
longer able to support the device for 25 minutes. The system will remain fully functional
while connected to AC power. When the battery replacement message appears on the
screen, contact your local Terumo Cardiovascular representative or call (800) 521-2818
for service.