C-1
Communication with Other Devices
The CDI OneView System has Modules that can connect to external devices to import
or export patient parameter values over the course of an operation.
When using a CDI OneView Module with the Terumo Advanced Perfusion
System 1 along with a CDI module, blood parameter values can also be displayed on
the Central Control Monitor.
Warning
Computer equipment in the operating room environment may
interfere with the operation of existing monitoring or therapeutic
devices and may be susceptible to interference from such
devices. To ensure that such interference will not occur, care
must be taken in the selection of computer equipment or printers
to be interfaced with the CDI OneView System and in the manner
in which this interface is accomplished.
Equipment connected to the Module’s port should be certied
according to IEC 60601-1 for medical equipment. All congurations
should be checked for combined system leakage current within
IEC 60601-1 leakage current specications.
Caution
Connecting unintended devices to the system could result in
unidentied risks to patients and/or users.
If unintended devices are connected to the system, the institution
should identify, evaluate and control any new risks that may arise.