Calibration and Circuit Setup
Sensors and Cuvettes
This section explains how to place the CDI Shunt Sensors and H/S Cuvette into the
extracorporeal circuit and how to attach Flow Sensors.
Prepare Disposables to Install Prepare Disposables to Install
p
To install Shunt Sensors, the following is required:
• CDI Shunt Sensor(s)
• CDI H/S Cuvette
• Access to the bypass circuit
For additional information, refer to disposable instructions for use.
Note: CDI Shunt Sensors and H/S Cuvettes can be used for either Arterial or Venous
application.
Warning
Once calibrated, do not remove and replace the sensor from the
BPM prior to use. Removal and replacement could affect the
measurement accuracy of the system.
Caution
Care should be taken to prevent BPMs from dropping onto
any hard surface or otherwise receiving severe shock. If a drop
occurs, the BPM should be carefully inspected for any damage
to the silver-domed thermistor contact, the optical pathways,
or the BPM housing surfaces. If the BPM had a sensor installed
and calibrated, you should replace the sensor and repeat the
calibration.