B-6
System Specifications
Environmental Specifications
Operating Conditions
Temperature 10 °C to 40 °C (50 °F to 104 °F)
Humidity 30% to 75%
Storage Temperatures
Hardware 10 °C to 40 °C (50 °F to 104 °F)
Shunt Sensors 0 °C to 35 °C (32 °F to 95 °F)
Calibration Gas 1 and 2 18 °C to 25 °C (64.4 °F to 77 °F)
Transport Temperatures
Hardware -20 °C to 60 °C (-4 °F to 140 °F)
Shunt Sensors 0 °C to 35 °C (32 °F to 95 °F)
Calibration Gas 1 and 2 -20 °C to 55 °C (-4 °F to 131 °F)
Storage Humidity
Hardware ≤ 75% (non-condensing)
Shunt Sensors 5% to 95% (non-condensing)
Calibration Gas 1 and 2 No limits
Transport Humidity
Hardware 10% to 95% (non-condensing)
Shunt Sensors 15% to 90% (non-condensing)
Calibration Gas 1 and 2 No limits
Atmospheric Pressure: 500 to 800 mmHg
Vibration/Shock: Proper performance of the system depends upon
high-delity optical alignment. Avoid vibration
and dropping components or cable-heads.