B-3
System Specifications
System Display Ranges
System Display Ranges are the full ranges in which parameter values will be displayed
by the system.
ParameterParameter Display RangeDisplay Range Resolution (*)Resolution (*)
pH 6.50 to 8.50 0.01
pCO
2
10 to 200 mmHg
1.3 to 26.7 kPa
1
0.1
pO
2
10 to 700 mmHg
1.3 to 93.3 kPa
1
0.1
K
+
1.0 to 9.9 mmol/L 0.1
Shunt Temp 1.0 to 45.0 °C 0.1
SO
2
35 to 100% 1%
HCT 12 to 45% 1%
Hgb 4.0 to 15.0 g/dL
40 to 150 g/L
0.1
Q
0 to 10.00 L/min
0 to 10,000 mL/min
0.001 < 1.0 L/min
0.01 ≥ 1.00 L/min
CI 0.1 to 10.0 L/min/m
2
0.1
BE -25 to 25 mEq/L 1
HCO
3
−
0 to 50 mEq/L 1
VO
2
1 to 400 mL/min 1
VO
2
i 1 to 1000 mL/min/m
2
1
DO
2
1 to 3000 mL/min 1
DO
2
i 1 to 1000 mL/min/m
2
1
rSO
2
match source device match source device
O
2
ER 0 to 100% 1%
SaSO
2
calc 35 to 100% 1%
SvSO
2
calc 35 to 100% 1%
(*) When no unit is listed, the same unit as listed in the Display range applies.