B-15
System Specifications
Caution
The ground equalization stud is used for external grounding
of the device. In some countries, regulations require potential
compensation. In this case, connect the cable from the potential
compensation network to the ground equalization stud. The
purpose of additional potential equalization is to equalize
potentials between different metal parts that can be touched
simultaneously or to reduce differences of potential which can
occur during operation between the bodies of medical electrical
devices and conductive parts of other objects. The ground
equalization stud on the Core is a 6 mm diameter pin conforming
to DIN 42801-part 1 Potential Equalization Leads - Connecting Pins.
Use a connecting lead with a 6 mm socket conforming to DIN
42801-part 2 Potential Equalization Leads - Connecting Sockets.
SecuritySecurity
The CDI OneView System incorporates several features to ensure the security of the
system. These features include inherent design elements as well as user settings and
functions.
General RecommendationsGeneral Recommendations
The CDI OneView System has been designed to utilize a proprietary communication
system to ensure only authorized CDI devices are used. As part of this implementation,
the system utilizes notications to the user of invalid values (shown as dashes “---“)
and error codes for improper communication as well as hardware issues. Action should
be taken in accordance with the Troubleshooting section in response to these events.
A USB drive can be utilized for transferring data to and from the system. In accordance
with your local policies, ensure the USB drive is properly secure. To protect the system
from threats originating from the USB drive, the CDI OneView Display restricts the
importing of case proles to ADMIN ONLY mode. Additionally, CDI software upgrades
are restricted to Terumo Cardiovascular Service personnel using Terumo Cardiovascular
authenticated USB drives.
Barcode scanning can be used for entering user passwords during login or for entering
K+ Code during BPM calibration. The following features are incorporated for barcode
scanning security:
The CDI Display will only accept barcodes of the specic predened format for K+
Code entry. If an incorrect format is used, the system will display an invalid barcode
error.
If scanning is not completed within 20 seconds, the system will display a barcode
read error.
User Accounts User Accounts
The CDI OneView Display includes User Management features for Administrator users
to create new users, assign password for user accounts, assign users a permission level
of admin or normal user, and congure Security settings.
Security settings should be carefully considered in accordance with the security needs
of the institution. Security settings include:
Passwords Complexity: Complexity requirements for passwords can be set to High,
Medium, or Low.