B-16
System Specifications
User Accounts (Continued)User Accounts (Continued)
• Login Requirements: User Access Control can be enabled to require all users to
login with an account.
Note: After 10 successive incorrect password attempts, the user account will be
locked for 30 minutes.
• Admin Inactivity Timeout: The Display can be congured to automatically logout
Admin users, mitigating threats that may arise from unattended system with admin
user logged in. The period of inactivity before the system logs the Admin user out
can be congured between 5 to 90 minutes.
Further details on user management and conguring Security settings are available in
Chapter 2, “Administrator Setup”.
Device Security, Event and Error LogsDevice Security, Event and Error Logs
The CDI OneView System records and maintains activity logs at various levels across
the CDI OneView components that may help with accountability and forensic purposes
in the event of security concerns. Device Logs includes below but not limited to:
• CDI OneView Display maintains security logs that contains user account access
information. The security logs can be saved as text les to the USB drive by
administrator users.
• CDI OneView Core logs the information on CDI devices connection status changes,
battery technical alarms and error logs.
• CDI OneView BPM Probe, H/S Probe and Flow maintains device conguration
change logs.
All CDI OneView devices maintain logs pertinent to self-test failures, memory integrity
checks and other service and maintenance purpose extended logs. These logs are
accessible by Terumo Cardiovascular Technical Support.
Ongoing Security InformationOngoing Security Information
In the event that new information regarding the security of the CDI OneView becomes
available, Terumo Cardiovascular will notify users of such information or required
actions. Such information includes security patches, identication of vulnerabilities, or
security lifecycle information.