4-9
Calibration and Circuit Setup
Sensors that have been calibrated successfully are ready for placement into the
extracorporeal circuit. Continue to the next page for further instructions.
The CDI OneView System remembers the most recent calibration data, even when
paused or switched off. The calibration data is replaced only when a new calibration is
performed.
Warning
Do not empty calibration uid from the sensor assembly, as
microsensors must be kept wet. If the Shunt Sensor is left
uncapped for more than a few minutes, dry-out may occur that
will affect sensor performance.