26
The visualisation of the schema of E
MPE
is identical to the one above, except that parameters
A and B are replaced by C and D. It is also possible that the technical specifications of certain
products indicate A = C and B = D.
The MICRO-HITE and MICRO-HITE+M are instruments with a “fixed zero”.
This means that in order to have measuring results in accordance with the
specifications indicated by the maximum permissible error, the reference
used in a measurement sequence has to be measured on the granite table
generally used for most applications.