1-2
- Do not place any heavy object on the power cord or squash the power cord.
Fire or electric shock may occur.
- Fully insert the power plug into the outlet. Faulty contact, allowing any metal to
contact the exposed terminal of the plug, or dust accumulated on the exposed
terminal of the plug may result in fire or electric shock.
- Conduct grounding work correctly. Otherwise, you may get an electric shock.
- Do not connect a device with data transmission specifications that are not
compatible. Fire or electric shock may occur. Contact Tomey Corporation or
our local distributor before using the instrument while connected to another
device.
- Install the instrument in a location not subject to direct sunlight, high
temperature and humidity, or air containing dust, salt, and/or sulfur. Otherwise,
failure or malfunction may occur.
- Install the instrument in a leveled stable location free of vibration or
mechanical impact. Otherwise, measurement cannot be conducted correctly.
Also, the instrument may topple over or fall down, resulting in fire or a serious
accident.
- Install the instrument in a location with ample clearance from other devices to
allow smooth inspection.
- Check the frequency, voltage, and allowable current (or power consumption)
of the power source.
- Check the power source (discharge condition, polarity, etc.).
- When the result of the axial length measurement unit is not stable, please use
the axial length value, anterior chamber depth, and crystal lens value which
are displayed in the axial length auxiliary measurement function only as
supplementary information to select adopted values, and do not use them for
IOL calculation or similar.
■
Precautions before operation
- Check the electrical contact of switches, polarity, dial setting, and meters, and
that the instrument is working correctly.
- Check that all cables are connected correctly.