3-131
3) Touch the active caliper line switching button (4) to select the caliper line to
be changed. The selected caliper line is displayed in red but the other lines
are displayed in yellow.
4) Touch the caliper line movement buttons (5) to change the caliper line
position. The changed measurement data is displayed in the edit data
display field (6) along with the movement of the caliper line.
The distance between the cornea caliper line (8) and the retina caliper line
(11) is displayed in Axial length, and the distance between the cornea
caliper line (8) and the caliper line for the front of the crystal lens (9) is
displayed in ACD, and the distance between the caliper line for the front of
the crystal lens (9) and the caliper line for the back of the crystal lens (10) is
displayed in Lens.
5) If you have edited it mistakenly, touch the “Initial Position” button (7) to
cancel the changes made, and the caliper line will return to the position
before the editing.
6) When you touch the “Apply” button (12), the changes made are applied,
and the measurement data displayed in the measurement data display field
(7) is overwritten with the edited measurement data, and the caliper screen
is closed. Touching the “Cancel” button (13) will discard the changes made,
and the caliper screen is closed.