5-5
■
b) Cleaning
■
Use absorbent cotton to clean the contact section of
the measurement probe. Otherwise, the contact
section may be damaged.
■
Holding and pulling the cord may damage the inner
core wires.
■
Do not clean the connecting section of the connector.
z When cleaning the main body of measurement probe, wipe it with
absorbent cotton dampened with antiseptic soap and water or alcohol,
and then with cotton dampened with water.
z When cleaning the tip of the measurement probe, wipe it with
absorbent cotton dampened with alcohol, and then with cotton
dampened with water.
z Wipe cords with a well wrung cloth dampened with water.
z Completely remove water after cleaning.
c) Disinfection/Sterilization
■
Effective chemicals vary depending on the types of
germs and viruses. Select appropriate chemicals and
immersion time based on the guidelines, articles,
literature, etc. provided by local administrative
organizations and/or societies.
■
Do not use EOG sterilizer for the eye cup.
■
Never use a high-pressure steam sterilizer.
■
Follow the instruction manual for the EOG sterilizing
conditions. Otherwise, safe sterilization cannot be
assured.