5-5
b) Cleaning
Use absorbent cotton to clean the contact section of
the measurement probe. Otherwise, the contact
section may be damaged.
Holding and pulling the cord may damage the inner
core wires.
Do not clean the connecting section of the connector.
z When cleaning the main body of measurement probe, wipe it with
absorbent cotton dampened with antiseptic soap and water or alcohol,
and then with cotton dampened with water.
z When cleaning the tip of the measurement probe, wipe it with
absorbent cotton dampened with alcohol, and then with cotton
dampened with water.
z Wipe cords with a well wrung cloth dampened with water.
z Completely remove water after cleaning.
c) Disinfection/Sterilization
Effective chemicals vary depending on the types of
germs and viruses. Select appropriate chemicals and
immersion time based on the guidelines, articles,
literature, etc. provided by local administrative
organizations and/or societies.
Do not use EOG sterilizer for the eye cup.
Never use a high-pressure steam sterilizer.
Follow the instruction manual for the EOG sterilizing
conditions. Otherwise, safe sterilization cannot be
assured.