EasyManuals Logo

Tomey UD-8000 User Manual

Tomey UD-8000
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #341 background imageLoading...
Page #341 background image
6-7
â– 
[B-mode image diagnosis]
z B-mode image is not displayed.
The B-mode probe is not connected correctly.
Check that the B-mode probe is firmly
connected.
z B-mode images are dark and not easy to see.
The setting of the total gain is low.
Turn TOTAL volume to increase the total gain.
z Ultrasound images are not clear or exact.
The B-mode probe is not connected correctly.
Check that the B-mode probe is firmly
connected.
Gel for ultrasound diagnosis is not used.
Apply gel for ultrasound diagnosis to the tip of
the probe. If the problem is not solved even
after the protective agent is applied, apply
more.
Noise is generated in the peripheral area
If there is any source of noise (devices such as
a motor, laser surgical equipment, etc.) near the
instrument, move it away from the instrument.
There are much multiechos.
Apply gel for ultrasound diagnosis to
sufficiently.