z “Probe error” is displayed when switching to the axial length measurement auxiliary
function.
The B-mode probe is not connected correctly.
Check that the B-mode probe is firmly
connected.
[Axial length measurement display function]
z Measurement cannot be performed in Auto mode.
The mode is set to Manual.
Set to auto measurement (Auto or Auto quick).
(Refer to “3.5.1 a) Setting the eye to be
measured, the axial length calculation method,
and the converted sonic velocity.”)
Noise is generated in the peripheral area.
If there is any source of noise (devices such as
a motor, laser surgical equipment, etc.) near
the instrument, move it away from the
instrument.
Patient’s sight is unstable.
Use the fixlight in the biometry probe or on the
chin rest to guide the sight of the patient.
The contact section of the biometry probe is
damaged.
If damaged, immediately stop measurement
and contact our local distributor.