EasyManuals Logo

Troxler 3440-L User Manual

Troxler 3440-L
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #121 background imageLoading...
Page #121 background image
Model 3440-L 7–11
7. EXTENDED FUNCTIONS
CALIBRATION STANDARD
The Calibration Standard function allows the operator or service
technician to input the factory calibration standard count.
To execute Calibration Standard function, press SHIFT and SPECIAL
for:
SPECIAL FUNCTION
YES- Next menu
1- STAT TEST
2- DRIFT TEST
Press 19 to display:
EXTENDED
FUNCTIONS
Enter Code ----
and Press ENTER
The Access Code must be entered. This code is for authorized personnel
only. If you need assistance with the code, contact your nearest Troxler
Representative.
Enter the code and press START/ENTER. The display will be:
YES- Next menu
1- Time/Date
2- Customer Name
3- Serial Number
Press YES three times for:
YES- Next menu
10- Calib. Std
11- Remote Cntrl
Press 1 0 to display:
Calib. Standard
Count: x
Input and
Press ENTER
Input the calibration standard count and press START/ENTER.
The display will return to the Extended Functions menu.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3440-L and is the answer not in the manual?

Troxler 3440-L Specifications

General IconGeneral
BrandTroxler
Model3440-L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals