Model 3440-L 7–9
7. EXTENDED FUNCTIONS
ERASE STANDARD COUNTS
The Erase Standard Counts function allows the operator or service
technician to remove all (four) standard counts from gauge memory.
To execute the Erase Standard Counts function, press 〈SHIFT〉 and
〈SPECIAL〉 for:
SPECIAL FUNCTION
YES- Next menu
1- STAT TEST
2- DRIFT TEST
Press 〈19〉 for the display:
EXTENDED
FUNCTIONS
Enter Code ----
and Press ENTER
The Access Code must be entered. This code is for authorized personnel
only. If you need assistance with the code, contact your nearest Troxler
Representative.
Enter the code and press 〈START/ENTER〉. The display will be:
YES- Next menu
1- Time/Date
2- Customer Name
3- Serial Number
Press 〈YES〉 two times for:
YES- Next menu
7- Erase Std Cts
8- Memory Clear
9- Burn-in
Press 〈7〉 for the display:
Press SHIFT/YES
to erase all
Standard Counts
in gauge?!
Press 〈SHIFT〉 and 〈YES〉 to erase all the standard counts. The display will
return to the Extended Functions menu.