6–6
PRINTING STAT TEST DATA
From the Select method of viewing... screen on page 6–4, press 2 for
the display:
Connect Printer
& press START
After connecting the printer to the serial connector on the gauge, press
START/ENTER for the following display. Figure 6–1 shows a sample
stat test printout.
- STAT TEST -
Downloading
data!
If the printer is not connected properly, an error message will be displayed.
********************************
Statistical Stability Test
********************************
Gauge Serial #: 18326
6/15/2007 7:21 AM
READING | COUNTS
# | DENSITY MOISTURE
1 2711 271
2 2740 271
3 2743 267
4 2742 274
5 2740 269
6 2758 262
7 2751 273
8 2758 265
9 2763 261
10 2746 268
11 2737 269
12 2764 267
13 2730 268
14 2751 270
15 2755 272
16 2731 262
17 2758 263
18 2750 269
19 2742 259
20 2710 270
Average: 2744 268
Ratio: 0.287 0.258
(PASS) (PASS)
Figure 6–1. Stat Test Printout Sample